|
Phone: 303/378-4577 |
Below are frequently asked questions (FAQs) sorted by category. Please contact us with any additional question, comment, or suggestion.
|
Category |
Question |
|
Can RenaiSys provide technical service and support for my DSR system? |
|
|
|
|
|
|
|
|
|
What does RenaiSys do during a preventative maintenance visit? |
|
|
|
|
|
|
|
|
Can RenaiSys provide technical service and support for my DSR system?
Yes. One of our members has been servicing and supporting the systems for many years.
The system does not initialize. What should I do?
Turn the system (including all computers) off. The following procedures are meant for you to try on you own risk. We shall not be held liable for the correctness of the procedure nor any damage to the system, loss of data, incidental damage or anything else. If at any point during the following procedure anything unexplained happens, please contact us or an other organization with the necessary expertise.
Manually position the carriages in the middle of their traveling ranges. Check that all cables are undamaged and properly connected. It often helps to disconnect cables and check the plugs and sockets. After that, reconnect the cables again. In addition, reseat the circuit boards in the electronics rack by pulling them out about an inch or two centimeters and pushing them back in, to make sure that they are seated firmly.
Turn on the power to the instrument. Do the lights come on? If not, check circuit breakers and fuses. If they are ok, a power supply could be broken. To verify this, turn the selector switch below the electronics rack and check the readings of the meter next to the switch. Warning, not everyone should be allowed to work inside the instrument due to the presence of electrical danger!
If the lights are on, do you notice anything unusual (for example a dark display, noise, or smoke) at this point? If not, double check the power supplies by turning the selector switch below the electronics rack and checking the readings of the meter next to the switch. From our experience, many problems, including intermittent problems, can be traced to faulty power supplies. So, please check the power supplies one more time!
If everything looks good so far, continue with the initialization by turning on the computer and entering the initialization command if needed. Watch the carriages very carefully. You might need to restart at the beginning of the procedure a couple times. Do all carriages start moving? Do they move in the proper direction? Do they move at the same speed? Once they hit the limits, do they reverse direction? Do they stop in the middle (even for a very short moment)? Do they move into the opposite limit? If one of the carriages moved into an opposite limit, stop monitoring, because anything from here on usually makes no sense and is only confusing.
If no carriage behaves any different than the others, a common problem could be suspected. Triple check the power sources. One of the unique circuit boards in the electronics rack, a computer problem, connection problem, or a software problem could be the culprit.
If one carriage behaves differently than the other three, a problem could be suspected with the one carriage system. Swapping cables to the drive systems could reveal a change. Checking the reference and end switches while the instrument is off, might point us in the right direction as well.
After a successful repair, the components need to be returned to their original location. Adjustments might be needed. A diagnostics program should be used to test the system. Calibration might be needed as well before it is used again in a production environment.
Can we replace a power supply ourselves?
First some warnings. The following procedures are meant for you to try on you own risk. We shall not be held liable for the correctness of the procedure nor any damage to the system, loss of data, incidental damage or anything else. Not everyone should be allowed to work inside the instrument due to the presence of electrical danger! If at any point during the following procedure anything unexplained happens, please contact us or an other organization with the necessary expertise.
Several organizations have successfully replaced the power supplies themselves. However, pre-adjustment and electronics adjustment have to be considered:
1. The power supply has to be carefully pre-adjusted to the voltages that are needed in the system. As they are sent out, the voltages might not be adjusted and might even damage the instrument (for example: 24V where there should be 15V).
2. Changing the power supply has an influence on the whole electronics of the instrument. Servo adjustments as well as other adjustments depend on the power supply. From experience, we have been able to replace the power supply without readjusting the rest of the electronics if the instruments were carefully installed and well maintained.
With a pre-adjusted power supply but without adjusting the rest of the electronics, most systems seem to work fine. Occasionally the systems "hang" and continue to work after a slight vibration (for example a carriage got touched). This could indicate the need to adjust the rest of the electronics. We recommend alerting a qualified technician during the next visit about the change of the power supply.
What does RenaiSys do during a preventative maintenance visit?
Below is a copy of our checklist:
1. Check with operator about backup, expectations,
and known problems
2. If software maintenance is included, check
on backup and install updates
3. Initialize system and run
diagnostics program for instrument and its peripherals
4. Check
optics (oculars, doves, zooms, base in/out, and measuring marks)
5.
Check illuminations
6. Check reference and end switches
7.
Check Digitizer, Printer, and other connected peripherals
8.
Repair any reported problem
9. Remove covers
10. Check
spindles, couplings, and moving cables
11. Check and if needed
tighten screws and nuts
12. Clean and lubricate bearings and
running surfaces
13. Clean optics, image carriers, and covers
14.
Check and if needed adjust power supply and electronics
15.
Install covers
16. Clean work area, monitor, keyboard, and other
external components
17. Initialize system and run diagnostics
program for instrument and its peripherals
18. Check calibration
and if needed recalibrate instrument
19. Ask operator to readjust
optics, illuminations, peripherals, check calibration, and to resume
work
20. Write service report
21. Review report with operator
to ensure correctness, completeness, and satisfaction
How much does it cost to move the system?
To answer this question as accurately as possible, we
would have to ask six additional questions:
1. Where is the system
currently located?
2. Where is the new location going to be?
3.
How about access at the two locations (are the doors as wide as the
complete system, is everything on the first floor, do we have to use
elevators or stairs)?
4. Should we include the help needed to lift
the heavy items?
5. Should we include the transportation and
packing materials?
6. Should we include insurance?
First, if you plan on moving the system within a building without having to de-assemble anything and without the need to tilt it, you might consider doing this without our presence. While keeping it leveled, the system can be carefully lifted and moved. After the move, we recommend to check the level and adjust the legs if needed. In addition, diagnostics programs make it possible to check the functions and the calibration of the instrument.
If our presence is requested, we typically apply the current time and materials rates. For a general estimate, we would calculate with one day (8 hours) work to dismantle the system and with two days to assemble, adjust, and calibrate the system. Depending on your answers to the questions, if you add travel cost and all the other costs, you end up with the total estimated cost for the move.
We have estimated costs for $2'000 and more. For a quotation that fits your requirements, please consider the questions and contact us.
Can you furnish the spare parts?
RenaiSys sells spare parts only if needed as part of a service. We focus on service and support. If part of a service, we forward any product (equipment, software, parts, ...) at the same cost and condition as they are sold to us. We stock no products. With good relations to the manufacturers, we use original parts when needed.
For DSR parts, we suggest you contact Vickie Acord at Leica (Vickie.Acord@gi.leica-geosystems.com) with the serial number of the instrument, part numbers if known, and the description of the parts.
What are the most commonly used spare parts?
|
Part Number |
Description |
Note |
|
LH580050 |
Freehand Cable X/Y |
|
|
LH580051 |
Freehand Cable Z |
|
|
LH581087 |
Freehand Switch Board |
Check LH581183 |
|
LH580070 |
Power Supply 28V |
|
|
LH581195 |
Power Supply 5V |
|
|
LH580085 |
Linear Encoder Short |
DSR18 only |
|
LH580086 |
Linear Encoder Long |
DSR18 only |
|
LH375350 |
Rotary Encoder 1000 Count |
|
|
LH581172 |
Micro-Switch |
|
|
LH587772 |
Rotary Switch |
For Panel Board |
|
LH581196 |
Motor |
|
|
LH581197 |
Motor with Gear |
|
|
|
|
|
|
LH581092 |
Heat Filter |
Check LH581064 |
|
LH585889 |
Heat Filter |
|
|
LH581170 |
Lamp 12V 100W |
|
|
LH585852 |
Lamp 12V 20W |
|
|
LH921479 |
Fan |
Check LH581198 |
|
|
|
|
|
LH581180 |
Panel Board |
|
|
LH580073 |
Tacho Board |
DSR18 only |
|
LH580078 |
SIF Board |
DSR18 only |
|
LH581178 |
SL2 Servo Logic Board |
Check LH580081 for DSR18 |
|
LH580084 |
Servo Amplifier Board |
DSR18 only |
|
LH581181 |
Amplifier |
Check LH581050 |
|
LH581058 |
DS1000 Control Board |
|
|
LH581083 |
Bus Interface Board |
|
|
LH581084 |
Data Buffer Board |
|
|
LH581085 |
Network Board |
Check LH581062 |
|
LH581176 |
P3 Control Board |
|
|
LH581199 |
Encoder Interface Board |
|
For copyright, trademark, and other legal information please read the linked pages. RenaiSys, LLC 1997 - 2008.